国产精品对白在线观看_亚洲av秘无码一区二区三麻豆_国产成人无码a∨制服丝袜综合_亚洲91久久久亚洲成人十八毛片ā级毛片_亚洲一区二区三区国产日韩精品_国产户外露出视频精品_免费国产一级黄片_久久综合日韩AV成人_b站在线免费进入网站_日本不卡在线播放

How to simplify maintenance of a HART pressure transmitter

How to simplify maintenance of a HART pressure transmitter

Complete maintenance of a HART transmitter involves three distinct operations: configuration,?calibration, and adjustment. Often configuration tasks can be accomplished only with a communicator, but adjustment and calibration require equipment that can accurately source pressure and measure current. Most communicators do not have the capability to measure or source analog values.

Configuring a HART transmitter consists of communicating with the transmitter and verifying its configuration data. It is beyond the scope of this application note to list the many different types of pressure transmitters and configurations, but common configuration parameters are listed below.

Calibrating a HART transmitter means to test and adjust its response so that the output accurately corresponds to the input throughout a specified range. To do this, an accurate source of pressure is required to apply a precisely known value to the transmitter. The output of the transmitter can then be compared to the known applied value, and an adjustment can be applied if needed. It is not possible to perform a calibration without comparing the transmitter’s response to a known, physical pressure.

Adjustment is the act of correcting the transmitter’s operation so that the output precisely matches the configured transfer function. In a common HART pressure transmitter there are several possible adjustment procedures:

  • Reranging: sets the lower and upper range points at required pressures, similar to adjusting zero and span on an analog transmitter.
  • Analog Output Trim: adjusts the transmitter’s analog output to match external standards. This essentially adjusts the transmitters digital-to-analog converter.
  • Zero Trim: adjusts the input sensor to compensate for pressure offsets, typically due to installation and orientation.
  • Sensor Trim: adjusts the factory sensor. This adjusts the analog-to-digital converter to compensate for sensor errors.

The adjustments that need to be performed on a HART transmitter will depend upon how the transmitter outputs are used. If only the 4-20 mA analog output signal is used, then Reranging, Analog Output Trim and Zero Trim are appropriate adjustments. If the transmitter’s digital outputs are used then Sensor Trim is necessary.

Get help