国产精品对白在线观看_亚洲av秘无码一区二区三麻豆_国产成人无码a∨制服丝袜综合_亚洲91久久久亚洲成人十八毛片ā级毛片_亚洲一区二区三区国产日韩精品_国产户外露出视频精品_免费国产一级黄片_久久综合日韩AV成人_b站在线免费进入网站_日本不卡在线播放

The 9144 Field Metrology Well has replaced the 9141 Dry-Well

The Fluke Calibration 9141 Dry-Well is now obsolete and has been replaced with the 9144 Field Metrology Well.

Top reasons to consider the 9144 Field Metrology Well over the 9141 Dry-Well

  • Better accuracy, stability, and radial uniformity
  • Dual heater zones improve vertical temperature gradient
  • Greater immersion depth
  • Modern user interface and industrial design
  • Process Electronics option to read external reference probes and UUTs

Comparing the 9141 Dry-Well with the 9144 Field Metrology Well (with and without Process Electronics)


?

9141 Dry-Well

9144 Field Metrology Well

9144-P Field Metrology Well with Process Electronics

Built-in thermometer readout

No

No

Yes

Temperature range

50 °C to 650 °C

50 °C to 660 °C

50 °C to 660 °C

Accuracy (at max)

± 1.0 °C

± 0.5 °C

± 0.07 °C

Stability (at max)

± 0.12 °C

± 0.05 °C

± 0.05 °C

Radial uniformity (at max)

± 0.12 °C

± 0.10 °C

± 0.10 °C

Axial uniformity (at max)

No spec

± 0 5 °C

± 0.5 °C

Heating speed (C/min)

52

41

41

Cooling speed (C/min)

22

17

17

Immersion depth (mm)

124

150

150

Learn more about the?9144 Field Metrology Well?